Ultra-High-Accuracy & Resolution Handheld Thermistor Thermometer
• Works with Standard OMEGA ® “400” Series Thermistor Probes with Phone Jack Termination
• Ultra-High Accuracy, up to ±0.015°C (±0.027°F)
• Includes NIST Traceable Calibration Certificate
• High Resolution, up to 0.01°F or °C
• -20 to 130°C (-4 to 266°F) Temperature Range
• Model HH42 has RS-232 Communications
• Hold Function, Auto Shutoff, Rate of Change Arrow Indicators, Over/ Under Range and Low Battery Indication
The Model HH41 handheld thermistor thermometer is a rugged, field tested unit that fits a wide variety of temperature measurement applications requiring ultra-high accuracy and high resolution at an economical price. The HH41 accepts OMEGA’s 400 Series thermistor probes with phone jack termination. The HH41 features display hold and °F/°C selection from the single front panel tactile feedback silicone function button. The unit’s LCD display features a low battery indication and also rate of temperature change arrows which, flashing at different speeds, inform the operator of the rate at which the temperature is changing. The unit has internal jumpers to disable selection of °F or °C from the front panel and enable display of either °F or °C only. Internal jumpers are also present to select either auto or manual shutoff mode. In the auto shutoff mode, the unit will automatically shut off 10 minutes after the last function key entry. Model HH42 has all the features of the HH41 with the addition of RS-232 communications. In situations where very high system accuracy is required, OMEGA offers a total system calibration service (probe plus meter); ask for CAL-4.
Specifications
Instrument Accuracy* vs. Temperature
Temp °C (°F) |
Accuracy °C (°F) |
-20 (-4) |
±0.016 (0.029) |
-10 (14) |
±0.015 (0.027) |
0 (32) |
±0.015 (0.027) |
10 (50) |
±0.015 (0.027) |
20 (68) |
±0.015 (0.027) |
30 (86) |
±0.015 (0.027) |
40 (104) |
±0.017 (0.031) |
50 (122) |
±0.019 (0.034) |
60 (140) |
±0.023 (0.041) |
70 (158) |
±0.029 (0.052) |
80 (176) |
±0.037 (0.067) |
90 (194) |
±0.049 (0.088) |
100 (212) |
±0.066 (0.119) |
110 (230) |
±0.088 (0.158) |
120 (248) |
±0.117 (0.211) |
130 (266) |
±0.156 (0.281) |
*Ambient temperature: 18 to 28°C (64 to 82°F) |
|
Resolution
Temp Range°C (°F) |
Resolution°C (°F) |
-20 to 102 (-4 to 215) |
102 to 130(215 to 266) |
0.01 (0.01) |
0.02 (0.05) |
Repeatability: 0.002 to 0.01°(C (-20 to 70°(C); 0.004 to 0.02°(C (-4 to 158°(F) typical for one week at constant ambient temperature Temperature Range: -20 to 130°(C (-4 to 266°(F) Reading Rate: 2 per second Display: 412 digit LCD Power: 9 Vdc battery, alkaline (included), optional 110 Vac or 220 Vac adaptor Battery Life: 20 hr, typical, alkaline battery Battery Indicator: Displays flashing battery indicator when less than 1 hour of battery life remains Temperature Unit Selection: °(F/°(C selectable from front panel function button, or internal jumpers to enable °(F or °(C only Shutoff Mode: Internal jumper selection for auto or manual shutoff mode; auto shutoff in 10 min with no function key entry Thermistor Sensor: Works with OMEGAR 400 Series thermistor (2252 ) probes with 14" phone jack (Model PPM-2) connection Operating Ambient: 10 to 40°(C (50 to 104°(F); 0 to 85% RH RS-232 Communications: 9600 baud, no parity Case Material: ABS Dimensions: 184.1 H x 83.8 W x 31.8 mm D (7.25 x 3.30 x 1.25") Weight: 312 g (11 oz)
|
|